Description
Advantages of the UC160-190?
- Reliable, Refrigerant-Free Thermoelectric Technology: With lifetimes exceeding 200,000 hours, the thermoelectric modules at the core of this chiller make for ultra-high intrinsic reliability
- Fits Inside Your Equipment: UC160-190 is less than one-sixth the size of its nearest competitor, measuring 7.5 x 5 x 7” (19 x 13 x 18 cm) and weighing only 8 lbs (3.5 kg)
- Precise Temperature Control: UC160-190’s digital PID control holds to ±0.1 °C through a range of 2 to 45 °C, even near ambient
- Easy Operation: Its standard, easy-to-use RS232 communications interface and digital controller for fixed temperature control or temperature cycling make operation of UC160-190 simple.
- Energy Efficient: Less than one-fourth the power usage of the nearest competitor (<200 W AC power)
- Air-cooled: Does not require house or facility chilled water. Can operate in any standard room, laboratory or office
- Lower Heat Load: Available in capacities from 160-190 Watts, well below the economical range of refrigerant-based chillers
- Versatile: Universal power input – so OEM’s need only one model for world-wide use. Multiple fluids: 27% propylene glycol solutions (preferred), ethylene glycol solutions
Specifications:
Operating Range: | 2 to 45 °C: (160/180W models) |
Ambient Temperature: | 10°C to 40°C non-condensing |
Stabillity/Repeatability: | ± 0.1°C with constant load (even near ambient) |
Cooling Capacity: | 160 to 190 Watts @ 20°C (20°C ambient) |
Noise (at 1 meter): | < 63 dBA |
Coolant / Process Fluid: | Koolance (27% propylene glycol / water mix) |
Process Fluid Fittings: | 1/8″ female CPC quick connect with shut-off valve |
Pump: | Magnetically-coupled gear pump with |
Tank Volume: | 75 ml with level sensor (optional sealable cap) |
Wetted Materials: | Aluminum + polymers or Copper + polymers |
Size (L x W x H): | 7.5″ x 5″ x 7″ (19 x 13 x 18 cm) |
Weight: | 8 lbs (3.5 kg) |
Power Input: | Universal: 100-240 VAC, 50/60 Hz, 2.8 amps max |
Power Consumption: | less than 200 Watts |
Operating Voltage: | 13.5 VDC, 15 amps max (universal input, laptop style power supply included) |
Communications: | Keypad or RS232 interface |
Alarms: | Temperature, fluid level, system or component failure (display, RS232 and dry contact) |
Standards: | TUV listed to UL, CAN/CSA and EN 61010-1, |
Warranty: | 1 year |